Power Meter Operation
2
E4418B Power Meter User’s Guide
43
Example
To make a measurement on channel A with a power sensor which has a
reference calibration factor of 99.8% and a calibration factor of 97.8% at the
measurement frequency.
•
Disconnect the power sensor from any source.
•
Press
.
•
Press
.
•
Press
. Use the
,
,
, and
keys to enter 99.8. Press
.
•
Connect the power sensor to the POWER REF output.
•
Press
.
•
Press
,
. Use the
,
,
, and
keys to
enter 97.8. Press
.
•
Connect the power sensor to the signal to be measured.
•
The measurement result is now displayed.
Zero
Ref CF
%
Cal
Cal Fac
%