Error Messages
4
User’s Guide
151
–230
Data corrupt or stale
This occurs when a
FETC?
is attempted and either a reset has been received or the power
meter state has changed such that the current measurement is invalidated (for example, a
change of frequency setting or triggering conditions).
–230
Data corrupt or stale;Please zero and calibrate Channel A
When
CAL[1|2]:RCAL
is set to
ON
and the sensor currently connected to channel A
has not been zeroed and calibrated, then any command which would normally return a
measurement result (for example
FETC?
,
READ?
or
MEAS?
) will generate this error
message.
–230
Data corrupt or stale;Please zero Channel A
When
CAL[1|2]:RCAL
is set to
ON
and the sensor currently connected to channel A
has not been zeroed, then any command which would normally return a measurement result
(for example
FETC?
,
READ?
or
MEAS?
) will generate this error message.
–230
Data corrupt or stale;Please calibrate Channel A
When
CAL[1|2]:RCAL
is set to
ON
and the sensor currently connected to channel A
has not been calibrated, then any command which would normally return a measurement
result (for example
FETC?
,
READ?
or
MEAS?
) will generate this error message
–231
Data questionable;CAL ERROR
Power meter calibration failed. The most likely cause is attempting to calibrate without
applying a 1 mW power to the power sensor.
–231
Data questionable;Input Overload
The power input to Channel A exceeds the power sensor’s maximum range.
–231
Data questionable;Lower window log error
This indicates that a difference measurement in the lower window has given a negative result
when the units of measurement were logarithmic.
–231
Data questionable;Upper window log error
This indicates that a difference measurement in the upper window has given a negative result
when the units of measurement were logarithmic.
–231
Data questionable;ZERO ERROR
Power meter zeroing failed. The most likely cause is attempting to zero when some power
signal is being applied to the power sensor.
–241
Hardware missing
The power meter is unable to execute the command because either no power sensor is
connected or it expects an Agilent E-Series or N8480 Series power sensor, and one is not
connected.
Summary of Contents for E4418B
Page 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Page 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...