2
Power Meter Operation
94
E4418B Power Meter User’s Guide
Self Test
The power meter has three distinct self test modes:
•
power on self test, which occurs automatically when you turn on the
power meter.
•
confidence check, which is accessed via the front panel and requires
your assistance in order to verify the POWER REF and measurement
path accuracy.
•
troubleshooting mode, which is accessed via the front panel or
remotely. The front panel softkey menu allows you to run individual
tests, whereas the remote command runs a complete series of tests as
listed in
Power On Self Test
The power on self test is performed automatically when the power meter
is turned on and takes approximately 10 seconds to complete. The power
on self test consists of the following tests:
•
RAM Battery
•
Calibrator
•
Measurement AssemblyFan
•
Serial Interface
•
Option 001 Battery
on page 98, if you require a description of
each individual test.
As the power on self test takes place, the message “Testing...” appears next
to the name of the test which is being performed. When each test is
complete, the message “Testing...” is replaced by either the message
“Passed” or “Failed”. The Option 001 Battery test can also produce the
message “Not Present”. If a failure occurs, the message “Power- up H/W
Err” appears. Any errors are also written to the error queue and can be
examined in the “Errors” screen by pressing
,
,
.
Error List
Summary of Contents for E4418B
Page 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Page 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...