![Agilent Technologies E4418B User Manual Download Page 173](http://html.mh-extra.com/html/agilent-technologies/e4418b/e4418b_user-manual_2868035173.webp)
Error Messages
4
User’s Guide
149
–123
Exponent too large
A numeric parameter was found whose exponent was larger than 32,000.
For example,
SENS:COUN 1E34000
.
–124
Too many digits
A numeric parameter was found whose mantissa contained more than 255 digits, excluding
leading zeros.
–128
Numeric data not allowed
A numeric value was received within a command which does not accept a numeric value.
For example,
MEM:CLE 24
.
–131
Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may have misspelled the suffix.
For example,
SENS:FREQ 200KZ
.
–134
Suffix too long
A suffix used contained more than 12 characters.
For example,
SENS:FREQ 2MHZZZZZZZZZZZ
.
–138
Suffix not allowed
A suffix was received following a numeric parameter which does not accept a suffix.
For example,
INIT:CONT 0Hz
.
–148
Character data not allowed
A discrete parameter was received but a character string or a numeric parameter was
expected. Check the list of parameters to verify that you have used a valid parameter type.
For example,
MEM:CLE CUSTOM_1
.
–151
Invalid string data
An invalid string was received. Check to see if you have enclosed the character string in single
or double quotes.
For example,
MEM:CLE “CUSTOM_1
.
–158
String data not allowed
A character string was received but is not allowed for the command. Check the list of
parameters to verify that you have used a valid parameter type.
For example,
LIM:STAT ‘ON’
.
–161
Invalid block data
A block data element was expected but was invalid for some reason. For example,
*DDT #15FET
. The 5 in the string indicates that 5 characters should follow, whereas in
this example there are only 3.
–168
Block data not allowed
A legal block data element was encountered but not allowed by the power meter at this point.
For example
SYST:LANG #15FETC?
.
Summary of Contents for E4418B
Page 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Page 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...