![Agilent Technologies E4418B User Manual Download Page 64](http://html.mh-extra.com/html/agilent-technologies/e4418b/e4418b_user-manual_2868035064.webp)
2
Power Meter Operation
40
E4418B Power Meter User’s Guide
•
Use
or
to move to other digits.
8
To confirm your choice, press the appropriate frequency units.
9
Connect the power sensor to the signal to be measured.
10
The measurement result is now displayed.
Example
To make a measurement using an Agilent E- Series power sensor. The
frequency of the signal to be measured is 100 MHz.
•
Disconnect the power sensor from any source.
•
Press
.
•
Press
.
•
Connect the power sensor to the POWER REF output.
•
Press
.
•
Press
,
. Use the
,
,
, and
keys to
enter 100. Press
.
•
Connect the power sensor to the signal to be measured.
•
The measurement result is now displayed.
Zero
Cal
Freq
MHz