Power Meter Operation
2
E4418B Power Meter User’s Guide
35
Zeroing and Calibrating Using TTL Inputs
You can use the TTL inputs on the rear panel Rmt I/O port to initiate
zero and calibration cycles on the power meter. The connector is an RJ- 45
Series shielded modular jack with the TTL input pins connected as shown
in
.
Figure 2-3
Rmt I/O Port TTL Inputs
The TTL inputs are active low and control the zero and calibration functions
as shown in
Pin Number
Connection
1
none
2
Ground
3
Upper Window TTL Output
4
Lower Window TTL Output
5
TTL Input 1
6
TTL Input 2
7
Ground
8
Ground
1 2 3 4 5 6 7 8
Table 2-2
TTL Inputs Control Logic
Input 1
Input 2
Operation
1
1
None
1
0
CAL
0
1
ZERO
0
0
CAL