![Agilent Technologies E4418B User Manual Download Page 189](http://html.mh-extra.com/html/agilent-technologies/e4418b/e4418b_user-manual_2868035189.webp)
Specifications
5
E4418B Power Meter User’s Guide
165
Settling Time
0 to 99% settled readings over the GPIB.
For Agilent 8480 Series power sensors
Manual filter, 10 dB decreasing power step:
Table 5-4
Settling Time
Auto filter, default resolution, 10 dB decreasing power step,
normal and x2 speed modes:
Number of Averages
1
2
4
8
16
32
64
128
256
512
1024
Settling Time (s)
(Normal Mode)
0.15
0.2
0.3
0.5
1.1
1.9
3.4
6.6
13
27
57
Response Time (s)
(x2 mode)
0.15
0.18
0.22
0.35
0.55
1.1
1.9
3.5
6.9
14.5
33
10 dB
10 dB
10 dB
20 dB
Minimum dBm
Power
150 ms
6.6 s
500 ms
200 ms
Settling
Times
Sensor
Dynamic
Range
Maximum dBm
10 dB
10 dB
10 dB
20 dB
Power
150 ms
3.5 s
350 ms
180 ms
Settling
Times
Sensor
Dynamic
Range
Maximum dBm
Minimum dBm
Normal Mode
x2 Mode
Typical
Typical
Summary of Contents for E4418B
Page 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Page 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...