Checklists
Hardware Installation
51
❏
Checkout Tune performed
❏
Tune Verification performed
❏
Sensitivity Check verification of EI Sensitivity performance
Switch CI MSD to CI operating mode (if applicable)
❏
MSD vented
❏
EI ion source removed
❏
CI ion source installed
❏
CI interface tip seal installed
❏
MSD pumped down and baked out for 2 hours
❏
Software set up for CI operation
❏
Purge Cal Valve performed
❏
Methane reagent gas flow set up
❏
Acceptable air and water background verified
PCI performance verification (if applicable)
❏
PCI autotune performed
❏
PCI sensitivity verified
NCI performance verification (if applicable)
❏
NCI autotune performed
❏
NCI sensitivity verified
Instrument Log Book Updated
❏
MSD serial number recorded on S/N report