Hardware Installation
Installation
Connecting the Helium Lines to the GC
1
Connect the 6-foot copper tubing (or a custom length) to the
carrier gas supply line (
).
2
Turn on the carrier gas flow at a low pressure, 5 to 8 psi
(35 to 55 KPa).
3
Let the line purge for 5 to 10 minutes.
4
Connect the other end of the tubing to the input of the gas
purifier.
5
Connect one end of the another length of copper tubing to
the output of the gas purifier.
6
Let the line and gas purifier purge for 5 to 10 minutes.
Figure 4
Connect helium
Copper supply tubing
Main supply On/Off valve
Helium carrier gas supply
Two-stage regulator
On/Off valve
Helium gas purifier (be sure the arrow
points in the direction of gas flow)
Copper supply tubing
C A U T I O N
Only remove the caps in the gas stream or the gas purifier will be
contaminated by air. This will ruin the gas purifier.