38
Hardware Installation
Installation
5
Connect the MSD power cord to an appropriate AC outlet.
6
Loosen the sideplate thumbscrews completely (shown in
) and open the analyzer.
7
Remove the interface column nut and blank ferrule from the
end of the GC/MSD interface.
8
Install the column in the GC/MSD interface. See the 5973N or
5973 inert MSD Hardware Manual for more information.
Figure 8
Loosen sideplate thumbscrews
Rear sideplate thumbscrew
(loosen completely)
C A U T I O N
Opening the analyzer with either sideplate thumbscrew tight will
irreparably damage the baseplate. Damage of this type is not covered
by the warranty.