Installation
Hardware Installation
39
9
Turn the GC on.
10
Set the GC column flow to 1.0 mL/min with constant flow
enabled.
11
Pump down the MSD.
Press on the side plate of the MSD to achieve a good seal.
Verify that the foreline pump and the front fan turn on, and
that the foreline pump stops gurgling within one minute.
12
After 10 minutes of pumping, turn off the MSD.
13
Turn off the GC temperature zones.
14
Set column flow to 1.0 mL/min with constant flow enabled.
15
Reinstall the MSD analyzer cover.
Figure 9
Rear panel view
Transmit (Tx) and receive (Rx) LEDs
Foreline pump power cord
Power cord receptacle
LAN cable
Remote start cable
Power (PWR) and heartbeat (heart) LEDs