Installation
Hardware Installation
33
8
Remove the cap from the diffusion pump (diffusion pump
MSDs only.)
9
Connect the foreline pump power cord to the receptacle on
the rear of the MSD.
C A U T I O N
Turning on the power to the MSD with the diffusion pump cap installed
will result in serious damage to the instrument. See the instructions
taped to the MSD.
WA R N I N G
The foreline pump must be plugged into the receptacle provided in
the back of the MSD in order to have adequate overcurrent
protection. Plugging the pump into a wall outlet will void the
warranty on the pump.
Figure 5
Preparing the vacuum system
Foreline hose (from high
vacuum pump outlet)
Exhaust outlet
(use 11-mm id hose)
Oil level window (sight glass)
Oil drip tray