36
Hardware Installation
Installation
6
Connect the second piece of stainless-steel tubing to the
outlet of the gas purifier.
7
Let the lines purge into a fume hood for 5 to 10 minutes.
8
Connect the other end of the tubing to the Gas A fitting on
the flow module as shown in
9
If another reagent gas is to be used, use more stainless steel
tubing to connect it to the Gas B fitting.
If you are going to connect isobutane, you will need another
gas purifier for the Gas B line. Observe the manufacturer’s
suggested pressure settings.
C A U T I O N
Do not connect isobutane without a gas purifier. Do not use the gas
purifier with ammonia.
Figure 6
Connecting the methane line
Gas B (alternate
reagent gas) line
Gas A (methane) line