Installation
Hardware Installation
43
Verifying CI System Performance
Special procedure for CI MSD
Refer to the
5973 inert MSD Hardware Manual
.
Verify PCI performance
1
Perform a positive CI autotune. See the 5973 inert MSD
Hardware Manual.
2
Verify positive CI performance. See the 5973 inert MSD
Hardware Manual.
Verify that rms signal-to-noise ratio meets the published
specification.
Verify NCI performance
1
Perform a negative CI autotune.
2
Verify negative CI performance.
Verify that rms signal-to-noise ratio meets the published
specification. See Agilent publication 5988-9991EN.
N O T E
There are no criteria for a “passing” CI autotune. If the CI autotune
completes, it passes. EMV at or above 2600 V, however, indicates a
problem with the system, and will cause problems meeting the sensitivity
specification.