28
Hardware Installation
Installation
Configuring the GC
1
If there is an ALS, install it now.
Follow the instructions provided with the ALS.
2
Install the Merlin Microseal (optional). Follow the
instructions provided with the Merlin Microseal.
3
Connect the GC power cord to the appropriate electrical
outlet.
4
Turn on the GC and verify that the GC passes self-test.
Keep the GC oven turned off.
5
Configure the column parameters for the MSD column that is
installed in the GC.
On the GC keypad, select
[Config
]
[Col 1]
and set the following
column parameters.
6
Increase the supply pressure of the carrier gas to
60 to 80 psi.
7
Set the column flow rate to 5 mL/min and split flow to
200 mL/min.
C A U T I O N
Do not use syringes with tapered needles with a Microseal.
Length:
30 M
ID:
0.25
Type:
HP-5 MS
Part number:
19091S-433