![Agilent Technologies 54657A User Manual Download Page 68](http://html.mh-extra.com/html/agilent-technologies/54657a/54657a_user-manual_2868526068.webp)
To save or recall traces
With the Measurement/Storage Module installed, the two volatile pixel
memories are replaced with four high-speed non-volatile memories.
In addition, 64 Kbytes of nonvolatile trace memory with data compression is
also provided. A data compression algorithm maximizes the number of traces
and front-panel setups that can be stored into this memory.
The total number of traces that can be saved depends on the complexity of
each trace. At least 4 highly complex traces, or up to 96 simple traces, can
be saved in this 64 Kbytes of memory. Storage time is less than 10 seconds.
1
Adjust Oscilloscope controls for desired stable display.
2
Press
Trace
.
A softkey menu with six choices appears. All of them are related to the trace
memories of the module.
•
Trace
Selects the trace memory (ALL, 1 to 100). When ALL is selected,
four additional softkeys are displayed.
a
All Off
Used to turn all 100 trace memories to off.
b
All On
Used to turn all 100 trace memories to on.
c
Clear All
Used to clear the contents of all 100 trace memories at
one time.
d
Review Traces
Used to view the contents of all trace memories
(with stored data) one at a time. View time is approximately 3
seconds. When selected, two additional softkeys are displayed.
Pause Review
pauses review cycle on current trace until key is
pressed again, and
Cancel Review cancels
the review cycle.
•
Trace X Off On
Turns the selected memory number to on or off. When on
is selected, trace status is temporarily displayed on the screen.
•
Save to Trace X
Saves the currently displayed waveform and front-panel
setup to the selected trace memory location.
•
Clear Trace X
Erases a previously stored waveform and front-panel setup
from the selected trace memory location.
Operating the Measurement/Storage Module
To save or recall traces
2–52
Summary of Contents for 54657A
Page 5: ...Installation 1 Reference Information 3 Operating the Measurement Storage Module 2 Index v ...
Page 6: ...vi ...
Page 8: ...3 Reference Information Operating Characteristics 3 3 Index Contents Contents 2 ...
Page 9: ...1 Installation ...
Page 16: ...1 8 ...
Page 17: ...2 Operating the Measurement Storage Module ...
Page 73: ...3 Reference Information ...
Page 80: ...3 8 ...
Page 87: ...Voltage Time Cursors Press this key to obtain this menu ...
Page 88: ...Print Utility Trace Trace Press this key to obtain this menu ...