Increment
This softkey is displayed only when
Save To Trace
is selected.
When On, all test violations are saved by incrementing the trace
number. The starting trace number is the one that is currently
selected. When the 64K compressed memory is full, the oldest
trace memory is overwritten, and the trace count continues
incrementing. When the trace count reaches 100, the number
resets to 1 (wraps around).
When Off, only the last test violation is saved, as test failure data is
written over previously stored data.
•
Previous Menu
Returns you to the previous softkey menu.
5
Toggle the softkeys to select the desired testing options.
6
Press the
Previous Menu
softkey.
7
Press the
Mask Test
softkey until
On
is selected.
Selecting on immediately starts the test using the test options specified. Test
indications are displayed on the display line as follows.
•
Pass
Indicates the displayed waveform passed the test.
•
Fail
Indicates the displayed waveform failed the test. Further testing,
and disposition of the failed data is dependent on the testing options
selected.
•
Acquisitions
Indicates the total acquisitions made during the test.
•
Failures
Indicates the total number (and percentage) of test failures that
occurred during the test.
Waveform Monitoring Operating Hint
Mask template testing can only be used in the Main Horizontal Mode, and when
Functions 1 and 2 are set to off.
The trace review softkey can be used to review all saved failures. See "To Save
or Recall Traces" in this chapter for more information.
Operating the Measurement/Storage Module
To start waveform monitoring
2–39
Summary of Contents for 54657A
Page 5: ...Installation 1 Reference Information 3 Operating the Measurement Storage Module 2 Index v ...
Page 6: ...vi ...
Page 8: ...3 Reference Information Operating Characteristics 3 3 Index Contents Contents 2 ...
Page 9: ...1 Installation ...
Page 16: ...1 8 ...
Page 17: ...2 Operating the Measurement Storage Module ...
Page 73: ...3 Reference Information ...
Page 80: ...3 8 ...
Page 87: ...Voltage Time Cursors Press this key to obtain this menu ...
Page 88: ...Print Utility Trace Trace Press this key to obtain this menu ...