![Agilent Technologies 54657A User Manual Download Page 62](http://html.mh-extra.com/html/agilent-technologies/54657a/54657a_user-manual_2868526062.webp)
Creating an overshoot testing mask
There are two parameters associated with the overshoot of a signal: the
percentage of overshoot and the settling time of the overshoot. A mask
template can be created to test the upper limit of these two parameters at
the same time. The following figure shows an example of a mask template for
testing overshoot.
Example of a mask template for testing overshoot
Figure 2– 15
Operating the Measurement/Storage Module
Creating an overshoot testing mask
2–46