Saving Test Violation Data Operating Hint
When Increment On is selected, traces
≥
3 to 100 are stored in the compressed
state. During the compression and storage of data, new signals are not
acquired or tested. The time it takes to compress and store data is less than 10
seconds.
When Increment On is selected, and multiple violations are desired, the On Fail
softkey must be set to Run (in the Test Options menu). The starting trace number
is the one that is currently selected. When the 64K compressed memory is full,
the oldest trace memory is overwritten, and the trace count continues
incrementing. When the trace count reaches 100, the number resets to 1 (wraps
around).
Operating the Measurement/Storage Module
To automatically save test violations
2–41