To edit the mask to test only a portion of a waveform
In certain circumstances, not all the points on the waveform need to be
tested. Only the area of interest needs to be tested. For example, to test the
amount of overshoot of a pulse, you only need to test the portion of the
waveform after the rising edge. You can select the test region by editing the
shape of the mask template.
1
Press
±
.
2
Press the
Mask Test
softkey.
3
Press the
Use Mask
softkey to select the desired mask number (1 or 2).
4
Press the
Define Mask Editor
softkey.
5
Toggle the
Edit Line
softkey to select the limit line that you want to
edit.
6
Turn the Delay knob to move the cursor to the starting location that
you do not want to test.
7
Turn the knob closest to the
Cursors
key to move the voltage
cursor until it reads "Don’t Care".
8
Press the
Mark
softkey.
9
Turn the Delay knob to move the cursor to the ending location of the
region that you do not want to test.
10
Turn the knob closest to the
Cursors
key to move the voltage
cursor until it reads "Don’t Care".
11
Press the
Connect
softkey.
This region of this particular limit line is not tested during the mask testing.
Operating the Measurement/Storage Module
To edit the mask to test only a portion of a waveform
2–36
Summary of Contents for 54657A
Page 5: ...Installation 1 Reference Information 3 Operating the Measurement Storage Module 2 Index v ...
Page 6: ...vi ...
Page 8: ...3 Reference Information Operating Characteristics 3 3 Index Contents Contents 2 ...
Page 9: ...1 Installation ...
Page 16: ...1 8 ...
Page 17: ...2 Operating the Measurement Storage Module ...
Page 73: ...3 Reference Information ...
Page 80: ...3 8 ...
Page 87: ...Voltage Time Cursors Press this key to obtain this menu ...
Page 88: ...Print Utility Trace Trace Press this key to obtain this menu ...