Function 2
Function 2 will plot differential or integral waveforms, or perform an FFT
using the input signals connected to the vertical inputs (1 and 2), or using
the function 1 waveform.
1
Press
±
.
2
Toggle the
Function 2 On Off
softkey to enable math function number 2.
3
Press the
Function 2 Menu
softkey.
4
Toggle the
Operand
softkey until the desired source is selected.
F1 uses the result waveform in function 1.
5
Press the
Operation
softkey until the desired operation is selected.
Results (F2) are displayed on the screen.
•
dV/dt
(differentiate) plots the derivative of the selected source using the
"Central Difference" formula. Equation is as follows:
d
n
+
i
=
c
n
–
c
n
+
1
+
2
i
∆
t
(
2
i
+
1
)
Where
d = differential waveform
c = input 1, 2, or function 1
i = data point step size
∆
t = point-to-point time difference
•
∫
dt
(integrate) plots the integral of the source using the "Trapezoidal Rule".
Equation is as follows:
I
n
=
∆
t
2
∑
(
c
n
+
c
n
+
1
)
Where
∆
t = point-to-point time difference
c = input 1, 2, or function 1
Operating the Measurement/Storage Module
Function 2
2–5
Summary of Contents for 54657A
Page 5: ...Installation 1 Reference Information 3 Operating the Measurement Storage Module 2 Index v ...
Page 6: ...vi ...
Page 8: ...3 Reference Information Operating Characteristics 3 3 Index Contents Contents 2 ...
Page 9: ...1 Installation ...
Page 16: ...1 8 ...
Page 17: ...2 Operating the Measurement Storage Module ...
Page 73: ...3 Reference Information ...
Page 80: ...3 8 ...
Page 87: ...Voltage Time Cursors Press this key to obtain this menu ...
Page 88: ...Print Utility Trace Trace Press this key to obtain this menu ...