The critical factors for creating the mask template are:
•
The vertical window of the middle region of the mask template determines
the upper limit of the overshoot.
•
The horizontal window of the middle region determines the upper limit of
the settling time.
•
The vertical window of the rightmost region determines the settling
window. Normally, the settling window is
±
5% or
±
10% of the V top
voltage.
Operating the Measurement/Storage Module
Creating an overshoot testing mask
2–47