![Agilent Technologies 54657A User Manual Download Page 59](http://html.mh-extra.com/html/agilent-technologies/54657a/54657a_user-manual_2868526059.webp)
The following procedure can be used to setup a mask template for testing
channel to channel delay.
In the oscilloscope setup, the controls should be selected to display the start
edge (channel 1) as the first edge on the display, and the stop edge
(channel 2) as the last edge on the display. The trigger source should be set
to trigger from channel 1. The mask template can be created by using an
external signal source to generate the signals identical to the ones that are
going to be tested.
1
Connect the desired signals to the oscilloscope.
2
Set the signal source(s) to generate a waveform identical to the ones
that you are going to test.
3
Press
±
on the oscilloscope, then press the
Mask Test
softkey.
4
Create a mask (at the desired tolerance) using Automask. Press
Previous Menu
softkey when finished.
Refer to "Create a Mask Template Using Automask" for more information on
using automask.
5
Press the
Define Mask Editor
softkey.
6
Toggle the
Edit Line
softkey to select the Min limit line. Use the
Mark
and
Connect
softkeys to edit the minimum line so only the last edge is
present (refer to previous figure).
Refer to "Create or Edit a Mask Using Line Segments" for additional
information.
7
Toggle the
Edit Line
softkey to select the Max limit line. Use the
Mark
and
Connect
softkeys to edit the minimum line so only the first edge is
present (refer to previous figure). Press
Previous Menu
softkey when
finished.
Operating the Measurement/Storage Module
Creating a delay testing mask
2–43
Summary of Contents for 54657A
Page 5: ...Installation 1 Reference Information 3 Operating the Measurement Storage Module 2 Index v ...
Page 6: ...vi ...
Page 8: ...3 Reference Information Operating Characteristics 3 3 Index Contents Contents 2 ...
Page 9: ...1 Installation ...
Page 16: ...1 8 ...
Page 17: ...2 Operating the Measurement Storage Module ...
Page 73: ...3 Reference Information ...
Page 80: ...3 8 ...
Page 87: ...Voltage Time Cursors Press this key to obtain this menu ...
Page 88: ...Print Utility Trace Trace Press this key to obtain this menu ...