Spectral Leakage
The FFT operation assumes that the time record
repeats. Unless there is an integral number of cycles of the sampled
waveform in the record, a discontinuity is created at the end of the
record. This is referred to as leakage. In order to minimize spectral
leakage, windows that approach zero smoothly at the beginning and end
of the signal are employed as filters to the FFT. The
Measurement/Storage Module provides four windows: rectangular,
exponential, hanning, and flattop. For more information on leakage, see
Agilent Application Note 243, "The Fundamentals of Signal Analysis"
(Agilent part number 5952-8898.)
FFT Operation
1
Press
±
.
2
Toggle the
Function 2 On Off
softkey to enable math function number 2.
3
Press the
Function 2 Menu
softkey.
4
Toggle the
Operand
softkey until the desired source is selected.
F1 uses the result waveform in function 1.
5
Press the
Operation
softkey until FFT is selected. Results (F2) are
displayed on the screen.
6
Press the
Units/div
softkey and rotate the knob closest to the
Cursors
key to set the vertical sensitivity of the resulting
waveform.
7
Press the
Ref Levl
softkey and rotate the knob closest to the
Cursors
key to set the reference level (top graticule line) of
the resulting waveform.
The
Autoscale FFT
softkey will automatically set Units/div and Ref Levl to bring
the FFT data on screen. Frequency Span is set to maximum. Steps 6 and 7 could
be replaced to say:
6
Press
FFT Menu
softkey.
7
Press
Autoscale FFT
softkey. Rotate Time/Div knob until freq span is around
the frequencies of interest.
8
Press the
FFT Menu
softkey.
A softkey menu with six softkey choices appears. Five of them are related to
FFT.
Operating the Measurement/Storage Module
FFT Measurement
2–10
Summary of Contents for 54657A
Page 5: ...Installation 1 Reference Information 3 Operating the Measurement Storage Module 2 Index v ...
Page 6: ...vi ...
Page 8: ...3 Reference Information Operating Characteristics 3 3 Index Contents Contents 2 ...
Page 9: ...1 Installation ...
Page 16: ...1 8 ...
Page 17: ...2 Operating the Measurement Storage Module ...
Page 73: ...3 Reference Information ...
Page 80: ...3 8 ...
Page 87: ...Voltage Time Cursors Press this key to obtain this menu ...
Page 88: ...Print Utility Trace Trace Press this key to obtain this menu ...