2D-LC User Guide
293
10
Troubleshooting and Diagnostics
Lab Advisor Service
&
Diagnostic
To upgrade/downgrade the module’s firmware carry out the following steps:
1
Download the required module firmware, the latest Lab Advisor software and
the documentation from the Agilent web.
http://www.agilent.com/en-us/firmwareDownload?whid=69761
2
For loading the firmware into the module
a
Select the folder on the hard drive where the Firmware package is stored.
b
Connect the Lab Advisor Software to your 2D-LC instrument.
Figure 153 Firmware Update
c
Press the
Lock
button.
The system is locked.
d
Select the required firmware version for the Resident and Main Firmware.
e
To update the firmware of the instrument, press the
Update
button.
This will require some time.
NOTE
Do not interrupt the power supply of the device and the PC during this procedure.