8
Run the System
Run the Experiment
2D-LC User Guide
208
2D Gradient:
Analysis 1.25 min, Equilibration 0.50 min
Analytical gradient - Shifted Gradient Shift 1D:
10 % B 0.00 min -
30 % B 20 min
60 % B 1.25 min
Flush gradient
not used
2
D Detector (DAD)
Diode-array
254 nm, Bandwidth 4 nm
Reference Wavelength 360 nm
Reference Bandwidth
100 nm
Peak width
80 Hz
Stop time
As pump/No limit
Table 19
Recommended conditions in 2D (HPLC) for multiple heart-cutting
Parameter
Value
Table 20
Recommended conditions in 2D (LC-MS)
Parameter
Value
Ion Source
Atmospheric pressure electrospray (Dual AJS ESI)
Ion Mode
Dual AJS ESI
Ion polarity
Both
Storage Mode
Both
LCMS Stream
MS
Acquisition Mode
Acquisition Mode MS1 Min Range (m/z) 50, Max Range (m/z) 1000, Scan
Rate (spectra/sec) 3.00
Instrument Parameters
Gas Temp
250 °C
Gas Flow
11 L/min
Nebulizer
40 psig
SheathGasTemp
350 °C
SheathGasFlow
12 L/min
Scan Segment
1
Scan Source Parameters
Vcap
3500 V