2D-LC User Guide
129
6
Method Parameters
Set the 2D-LC Method parameters
Define the
1
D Pump FLow
1
Set the
1
D Pump Flow
(range 0 – 5.0 mL/min).
This setting defines the flow in the first dimension being used while 2D-LC is
active.
Any changes of the Flow parameter in the 2D-LC UI are automatically
synchronized with the Method User Interface of the
1
D pump.
Figure 67
Method User Interface of the
1
D pump
Define the
2
D Pump Flow
1
Set the
2
D Pump Flow
(range 0 – 5.0 mL/min).
This setting defines the flow in the 2nd dimension being used while 2D-LC is
active (within
2
D time segments where mode is not equal to
Off
).
2
To set and use idle flow, select check box
Idle flow
.
The field to define the idle flow is active.
The setting in this field defines the flow in the 2nd dimension that is used
while the 2D-LC mode is
Off
(range 0 – 5.0 mL/min) and no cut is analyzed.
Figure 68
Interface for the flow settings of the
1
D pump and
2
D pump
NOTE
The selection of the solvents must be done in the standard pump method user
interface.
NOTE
If
Idle flow
is not selected, the
2
D Flow is also used when no
2
D analyses take
place.