10
Troubleshooting and Diagnostics
Lab Advisor Instrument Control
2D-LC User Guide
284
Instrument Control of the 2D-LC Cluster
1
Select
Instrument Control
of the
2
D pump (2D-LC cluster).
2
Change the settings of the
2
D pump as required.
3
To identify a valve, select the valve from the
Valve
drop-down list.
The following instrument setups are possible:
•
One 2D-LC valve
•
Three valves:
•
One 2D-LC valve
•
Two MHC valves
Figure 146 Example of a 2D-LC instrument with a 2D-LC ASM valve and two MHC valves
When
Control the behavior of the
2
D pump and the 2D-LC valves.
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
•
Documentation provided with the Agilent Lab Advisor online help
•
2D-LC Manual
Procedure to follow:
•
Close the current Acquisition client window
•
Close instrument connection from OpenLab Control Panel
NOTE
To use instrument control of the
2
D pump, the 2D-LC hardware license must be
active.