2D-LC User Guide
163
6
Method Parameters
Preview (2D-LC)
Figure 83
Chromatogram with missed peaks marked red
MHC cuts
(time based)
This function uses the continuous flow-through principle. The cuts are visualized as light green bars. The
dark line on the right edge of the bar indicates the switching time of the 2D-LC valve and the end of parking
the peak. Cuts can be marked and moved to another position in the preview window.
HiRes cuts
(time-based)
HiRes cuts (time-based)
are visualized and marked as light green bars. Depending of the peak width the cuts
can vary from 2 to maximal 10 cuts. Compared to MHC, HiRes cuts have two dark lines one on the left side
one of the ride side of the bar which reflects the switching before and after parking a peak. The left dark line
defines the start time of one high resolution sampling event.