7.3.2
RANGE OF RADIOGRAPHIC PARAMETERS
PARAMETER
RANGE
PARAMETER
RANGE
kVp
From 40 kVp 150 kVp in 1 kV steps.
mA
From 10 mA to 800 mA through the following mA stations:
10, 12.5, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800.
(Depending on the Generator model)
mAs
Product of mA x Time values from 0.1 mAs to 500 mAs
ms
From 1 to 10000 milliseconds through the following Time stations:
1, 2, 3, 4, 5, 6, 8, 10, 12, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640, 800,
1000, 1250, 1600, 2000, 2500, 3200, 4000, 5000. 6400, 8000 and 10000.
AEC
mAs: 0.1 mAs to 500 mAs
AEC
Exposure Time: Nominal shortest irradiation Time = 14 ms
Max ms
with DR Detector
550, 1000.
7.3.3
DUTY CYCLE
The Generator duty cycle is continuous, but limits should be set during
installation depending on the capacity of the X-ray Tube.