DX-D 300 U-Arm
User Manual
0171F EN 20181009
35
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY
(IEC 60601-1-2:2014)
The Radiographic Room DX-D300 is intended for use in an electromagnetic environment specified below.
The customer or Operator of thisRadiographic Room DX-D300 should assure that it is used in such an environment.
Immunity Test
IEC 60601-1-2:2014
Test Level
Compliance Level
Electromagnetic environment -
guidance
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
3 Vrms
from 80 MHz to 2.7 GHz
(80% AM at 1 kHz)
Refer to next table “IMMUNITY
3 Vrms
from 80 MHz to 2.7 GHz
(80% AM at 1 kHz)
Refer to next table “IMMUNITY
Proximity fields from RF
wireless Communications
equipment
IEC 61000-4-3
Refer to next table “IMMUNITY
REQUIREMENTS FOR RF
WIRELESS COMMUNICATIONS
EQUIPMENT”
Refer to next table “IMMUNITY
REQUIREMENTS FOR RF
WIRELESS COMMUNICATIONS
EQUIPMENT”
Portable RF communications equipment
(including peripherals such as antenna
cables and external antennas) should be
used no closer than 30 cm to any part of the
equipment, including cables specified by
manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Conducted disturbances
induced by RF fields
IEC 61000-4-6
3 Vrms
from 150 kHz to 80 Mhz
6 Vrms in ISM bands
from 150 kHz to 80 MHz
(80% AM at 1 kHz)
3 Vrms
from 150 kHz to 80 Mhz
6 Vrms in ISM bands
from 150 kHz to 80 MHz
(80% AM at 1 kHz)
performance of this equipment could result.
NOTE -- The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz;
26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz; 3.5 MHz to 4.0 MHz; 5.3 MHz to 5.4 MHz; 7 MHz to 7.3 MHz; 10.1 MHz
to 10.15 MHz; 14 MHz to 14.2 MHz; 18.07 MHz to 18.17 MHz; 21.0 MHz to 21.4 MHz; 24.89 MHz to 24.99 MHz; 28.0 MHz to 29.7 MHz; and 50.0 MHz
to 54.0 MHz.
Summary of Contents for DX-D 300
Page 1: ...DX D 300 U Arm User Manual Technical Publication 0171F EN 20181009 ...
Page 2: ......
Page 8: ...DX D 300 U Arm User Manual 0171F EN 20181009 iv ...
Page 56: ...DX D 300 U Arm User Manual 48 0171F EN 20181009 This page intentionally left blank ...
Page 100: ...DX D 300 U Arm User Manual 92 0171F EN 20181009 This page intentionally left blank ...
Page 103: ......