DX-D 300 U-Arm
User Manual
86
0171F EN 20181009
7.3
X-RAY GENERATOR
7.3.1
FACTORS
GENERATOR MODEL
SHF-515-1T-LS
RAPIDT 50 kW
SHF-515-1T-HS
RAPIDT 50 kW
SHF-535-1T-LS
RAPIDT 50 kW
SHF-535-1T-HS
RAPIDT 50 kW
SHF-635-1T-HS
RAPIDT 64 kW
SHF-835-1T-HS
RAPIDT 80 kW
Maximum Power kW
50 kW
64 kW
80 kW
Maximum mA
640 mA
800 mA
Maximum kVp
150 kVp
Power Line
A
C / D
C / D
D
Power Output
(@ 0.1 s)
640 mA @ 78 kVp
500 mA @ 100 kVp
400 mA @ 125 kVp
320 mA @ 150 kVp
640 mA @ 100 kVp
500 mA @ 128 kVp
400 mA @ 150 kVp
800 mA @ 100 kVp
640 mA @ 125 kVp
500 mA @ 150 kVp
Compatible
X-Ray Tubes
E7884X
E7252X
E7254FX
E7869XX
E7252X
E7254FX
E7869XX
E7884X
E7252X
E7254FX
E7869XX
E7252X
E7254FX
E7869XX
E7252X
E7254FX
E7869XX
E7254FX
E7869XX
POWER LINE AND NOMINAL WORKING FREQUENCY
A
C
D
230 / 240 VAC,
Single-Phase, 50 / 60 Hz
400 / 415 / 440 VAC,
Three-Phase, 50 / 60 Hz
480 VAC,
Three-Phase, 50 / 60 Hz
Line voltage automatic compensation:
±
10%.
Maximum line regulation for maximum kVA demand: 5%.
NOTES: -- For Generators operating with lines at 208 VAC or below an Auxiliary Boost Transformer is required to adequate the
line voltage to 230 / 240 VAC.
-- For 80 kW Generators operating with lines at 400 / 415 / 440 VAC an auxiliary boost transformer is required to adequate
the line voltage to 480 VAC (or 530 VAC).
Summary of Contents for DX-D 300
Page 1: ...DX D 300 U Arm User Manual Technical Publication 0171F EN 20181009 ...
Page 2: ......
Page 8: ...DX D 300 U Arm User Manual 0171F EN 20181009 iv ...
Page 56: ...DX D 300 U Arm User Manual 48 0171F EN 20181009 This page intentionally left blank ...
Page 100: ...DX D 300 U Arm User Manual 92 0171F EN 20181009 This page intentionally left blank ...
Page 103: ......