DX-D 300 U-Arm
User Manual
0171F EN 20181009
53
UNDERTABLE POSITION:
Press and hold this button to place automatically
the U-Arm in Undertable position. When the U-Arm arrives to Undertable
Position, the movement stops and the “
Safety Lock
” button lights up. Also
refer
to Section 4.1.3
THORAX POSITION:
Press and hold this button to place automatically the
U-Arm in Thorax position. When the U-Arm arrives to Thorax Position, the
movement stops and the “
Safety Lock
” button lights up. Also
refer to
Section 4.1.3.
UNDERTABLE AND THORAX POSITIONING ARE PROVIDED
WITH AUTOMATIC MOVEMENT OF THE U-ARM, CHECK
THAT NO PERSONS OR OBJECTS ARE WITHIN THE RANGE
OF THE U-ARM.
LOW / HIGH SPEED
: Movements are performed by default at High Speed.
Press this button (lights up in the Touch Screen Console) to perform
movements at Low Speed. Press again to deselect Low Speed.
SAFETY LOCK
: This button lights up in the Touch Screen Console when
Undertable or Thorax positioning is reached. Also when any other safety device
is activated, for example the DR Detector Assembly Safety Band or the U-Arm
Gauges. The Operator must press it before performing any other motion. The
Safety Lock stops any motion of the U-Arm. Also, this button resets all
recoverable errors.
DR DETECTOR ASSEMBLY ROTATION
: This motorized movement allows a
maximum of
+45
o
Clockwise movement and a minimum of -45
o
Counterclockwise movement. It includes a detent at 0
o
and also a detent when
the DR Detector Assembly is in parallel with respect to the floor. Press button
and release at desired position. The angle reached is shown in the Rotation
Display and in the Touch Screen.
The DR Detector Assembly Rotation is continuously indicated in the Touch
Screen Console (Detector Angle) and when the Detector has reached to an
Angle limit, a message appears on the Screen:
“SW LIMIT DET CCW”
or
“SW
LIMIT DET CW”.
Summary of Contents for DX-D 300
Page 1: ...DX D 300 U Arm User Manual Technical Publication 0171F EN 20181009 ...
Page 2: ......
Page 8: ...DX D 300 U Arm User Manual 0171F EN 20181009 iv ...
Page 56: ...DX D 300 U Arm User Manual 48 0171F EN 20181009 This page intentionally left blank ...
Page 100: ...DX D 300 U Arm User Manual 92 0171F EN 20181009 This page intentionally left blank ...
Page 103: ......