DX-D 300 U-Arm
User Manual
36
0171F EN 20181009
IMMUNITY REQUIREMENTS TO RF WIRELESS COMMUNICATIONS EQUIPMENT
(IEC 60601-1-2:2014)
The Radiographic Room DX-D300 is intended for use in an electromagnetic environment specified below.
The customer or Operator of this Radiographic Room DX-D300 should assure that it is used in such an environment.
Band
a)
(MHz)
Modulation
b)
Distance
(m)
Immunity Test Level
(V/m)
380 -- 390
Pulse modulation
b)
18 Hz
27
430 -- 470
FM
c)
¦
5 kHz deviation
1 kHz sine
28
704 -- 787
Pulse modulation
b)
217Hz
9
800 -- 960
Pulse modulation
b)
18Hz
0.3
28
1700 -- 1990
Pulse modulation
b)
217Hz
0.3
28
2400 -- 2570
Pulse modulation
b)
217Hz
28
5100 -- 5800
Pulse modulation
b)
217Hz
9
a)
For some services, only the uplink frequencies are included.
b)
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation,
it would be worst case.