GUIDANCE AND MANUFACTURER’S DECLARATION -- ELECTROMAGNETIC IMMUNITY
(IEC 60601-1-2:2007)
The
Radiographic
Room
DX-D300
is
intended
for
use
in
the
electromagnetic
environment
specified
below.
The
customer
or
the
user
of
this
Radiographic
Room
DX-D300
should
assure
that
it
is
used
in
such
an
environment.
Immunity Test
IEC 60601-1-2:2007
Test level
Compliance Level
Electromagnetic environment -- Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
¦
6kV contact
¦
8kV air
¦
6kV maximum
¦
8kV maximum
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical
fast
transient/burst
IEC 61000-4-4
¦
2kV
for power supply lines
¦
1kV
for input/output lines
¦
2kV
for power supply lines
¦
1kV
for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000--4--5
¦
1kV line(s) to line(s)
¦
2kV line(s) to earth
¦
1kV
symmetrical coupling
¦
2kV asymmetrical coupling
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines.
IEC 61000--4--11
< 5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in U
T
)
for 5s
>95%
for 0.5 periods
60%
for 5 periods
30%
>95%
for 250 periods
Mains
power
quality
should
be
that
of
a
typical
commercial
or
hospital
environment.
If
the
user
of
the
X-ray
Unit
requires
continued
operation
during
power
mains
interruptions,
it
is
recommended
that
the
X-ray
Unit
be
powered
from
a
uninterruptible
power
supply
or
a
battery
Power frequency (50/60
Hz) magnetic field
IEC 61000--4--8
3 A/m
3A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE -- U
T
is the a.c. mains voltage prior to application of the test level.
Summary of Contents for DX-D 300
Page 1: ...DX D 300 U Arm User Manual Technical Publication 0171F EN 20181009 ...
Page 2: ......
Page 8: ...DX D 300 U Arm User Manual 0171F EN 20181009 iv ...
Page 56: ...DX D 300 U Arm User Manual 48 0171F EN 20181009 This page intentionally left blank ...
Page 100: ...DX D 300 U Arm User Manual 92 0171F EN 20181009 This page intentionally left blank ...
Page 103: ......