SECTION 5
OPERATING SEQUENCES
5.1
START-UP ROUTINE
Start-up the system as described in Section 3.
5.2
X-RAY TUBE WARM-UP PROCEDURE
Before effecting X-ray exposures ensure that the Tube is
properly warmed-up. Make sure that no persons will be
inadvertently exposed to unnecessary X-rays during this
procedure.
Routine exposures should not be effected unless the Tube is previously
warmed-up, this prolongs X-ray Tube life.
It is recommended that the following procedure will be performed for X-ray Tube
warm-up, at the start of each day and when the Tube selected has not been in
use for approximately one hour.
This warm-up procedure is used for a typical X-ray Tube.
Consult the X-ray Tube manufacturer instructions for the
actual Tube in use, comparing its recommendations with
this procedure. If there is conflict with this procedure,
comply with the Tube manufacturer’s instructions.
Perform X-ray Tube warm-up as follows:
Close the collimator blades fully.
Select 70 kV, 100 mAs, 200 mA and 500 ms exposure.
Insure that no one will be exposed.
Make a total of three exposures, 15 seconds apart.
Excessive filament evaporation shortens X-ray Tube life.
Minimize evaporation by keeping Exposure “Preparation”
time to an absolute minimum.
Summary of Contents for DX-D 300
Page 1: ...DX D 300 U Arm User Manual Technical Publication 0171F EN 20181009 ...
Page 2: ......
Page 8: ...DX D 300 U Arm User Manual 0171F EN 20181009 iv ...
Page 56: ...DX D 300 U Arm User Manual 48 0171F EN 20181009 This page intentionally left blank ...
Page 100: ...DX D 300 U Arm User Manual 92 0171F EN 20181009 This page intentionally left blank ...
Page 103: ......