DX-D 300 U-Arm
User Manual
72
0171F EN 20181009
Table 4-1 (Cont.)
Error and Warning Messages
Error No DESCRIPTION (IN SCREEN)
WHAT TO DO
15
Movements locked
Release the Safety Lock Button
16
Move Button Released
Operator stopped pressing the Auto positioning button.
Press and hold until auto positioning has finished.
17
Incompatible Movements
A movement button has been pressed while auto
positioning was in progress. Release the movement button.
18
Opposite Move
An opposite movement button has been pressed while
other movement was in progress. Release the opposite
movement button.
19
Calibration Warning
Turn the Unit OFF/ON, if error remains, call field service.
20
Program not Selected
Move button has been pressed without having selected a
auto--position. Move button has been pressed again after
the target position has been reached.
U--arm position parameters are being changed during
movement.
21
Movement not enabled
Turn the Unit OFF/ON, if error remains, call field service.
22
Movement stop
The Unit has reached a position.
23
Inverters Powering up
Wait until the Unit has finished the starting sequence.