Anticrushing System: When the U-Arm is
moving down
or
when it rotates
,
it will always reverse the movement if the Anticrushing System is activated to
avoid something kept between the Detector Assembly and the floor.
1
2
3
1
2
3
When the U-Arm is
moving up
and the Anticrushing System is activated, it will
make the reverse movement or stop depending on the following situations:
The system only reverses down automatically when the Anticrushing
System is activated and the angle is between --30
o
and 50
o
(the angle
interval where the Detector Assembly could be under the table). In any
other case it stops.
1
2
3
--30
o
50
o
In the rest of the angle ranges (from 51
o
to 120
o
) and in horizontal
position, the system just stops.
51
o
120
o
In case the Unit remains continuously blocked, it can be moved by
pressing and holding the “Move” button and then pressing the
movement button required.