5.4
AEC OPERATION
The proper use of AEC requires accurate patient positioning. For examination
using AEC, the operator will need to select the desired AEC parameters as
follows:
1.
Make sure that the X-ray Tube is properly warmed-up.
2.
Position the patient for the examination.
3.
Select the “
DR Detector
” workstation and enter in AEC mode by
selecting at least one Area Detector “
Field
” on the Software Console.
4.
If required, choose another
“Sensitivity”
and adjust the “
Density
” setting
(“
0
” is the normal setting).
5.
Select the technique parameters using the Generator controls on the
Console.
6.
Continue with the radiographic operation.
(Refer to Section 5.3 -- step 4.)
5.4.1
HOW TO VERIFY THE PROPER FUNCTIONING OF THE AUTOMATIC EXPOSURE
CONTROL
This procedure is not mandatory, it is only a method so that the
operator can verify the proper functioning of the Automatic
Exposure Control.
1.
Ensure that X-ray Tube has been properly warmed up.
2.
Align and center the X-Ray Tube to the DR Detector.
3.
Set a SID of 1 m (40”).
4.
Collimate the X-Ray beam so that it completely covers all three Ion
Chambers (Left, Center and Right).
5.
Place on the Table-Top and within the X-Ray beam a homogeneous
phantom (e.g. a bucket with 10 cm of water) that covers all three Ion
Chambers.
6.
Set a technique, for example: 70 kVp, 250 mA, 1.0 second back-up time.
Note
.
Summary of Contents for DX-D 300
Page 1: ...DX D 300 U Arm User Manual Technical Publication 0171F EN 20181009 ...
Page 2: ......
Page 8: ...DX D 300 U Arm User Manual 0171F EN 20181009 iv ...
Page 56: ...DX D 300 U Arm User Manual 48 0171F EN 20181009 This page intentionally left blank ...
Page 100: ...DX D 300 U Arm User Manual 92 0171F EN 20181009 This page intentionally left blank ...
Page 103: ......