DX-D 300 U-Arm
User Manual
0171F EN 20181009
37
2.9
QUANTITATIVE INFORMATION
The following tables show the Quantitative Information associated
to this equipment according to the Standard IEC 60601--1--3:2008
+ AMD1:2013. These tables illustrate loading factors for image
performance and supply Dose indication examples. Therefore,
they are an example of the adjustment of Loading Factors, Focal
Spot Selection, SID and Collimator opening, which affect to the
radiation quality or to the radiation dose rate applied in normal use.
2.9.1
FUNCTIONAL TESTS PERFORMED TO OBTAIN THE QUANTITATIVE INFORMATION
Equipment:
Rad Positioner with Ralco Collimator.
Instrumentation used:
Dosimeter: Vacudap
Dosimeter: Unfors
Rectangular Phantom made of Polymethyl-methacrylate (PMMA)
layers: 25 cm x 25 cm x 20 cm.
Test Details:
Minimum SID:100 cm.
Maximum SID:180 cm.
Open Collimator size: 13 cm x 13 cm (min.), 43 cm x 43 cm (max.)
The measurements were made with the exposure parameters shown on
the results table:
KVp Range: 40 KVp, 60 KVp, 80 KVp, 100 KVp, 125 KVp
mAs Range: 1 mAs, 2 mAs, 10 mAs, 50 mAs, 100 mAs
Performed measurements of Air Kerma or Air Kerma Rate at the
following designated positions:
-- Distance SID doses
-- Patient (Phantom) Entrance doses and Entrance doses Rate
-- Patient (Phantom) Output doses and Output doses Rate
-- Collimator Output doses
Note
.
Summary of Contents for DX-D 300
Page 1: ...DX D 300 U Arm User Manual Technical Publication 0171F EN 20181009 ...
Page 2: ......
Page 8: ...DX D 300 U Arm User Manual 0171F EN 20181009 iv ...
Page 56: ...DX D 300 U Arm User Manual 48 0171F EN 20181009 This page intentionally left blank ...
Page 100: ...DX D 300 U Arm User Manual 92 0171F EN 20181009 This page intentionally left blank ...
Page 103: ......