DX-D 300 U-Arm
User Manual
68
0171F EN 20181009
4.5
TUBE-COLLIMATOR ASSEMBLY POSITIONING
Tube-Collimator Assembly is equipped with a blocking system (on top)
consisting of a lock-lever and a ratchet. In order to change angle position of the
Tube-Collimator Assembly in relation to its transverse axis, loose the lock-lever
and lift the ratchet. Ratchet has position detents every 45
o
.
As well, when the operator rotates the Tube-Collimator Assembly, the message
“Tube Rot”
appears in the U-Arm Software Console whenever the Detector
workstation is selected.
The Tube-Collimator Assembly can
be clockwise rotated up
to +1
80
o
and counterclockwise rotated up to --90
o
. Rotation
over these limits may damage the U-Arm cables.
180
o
90
o