Table 4-2 (Cont.)
Information Messages
MESSAGE ON SCREEN
WHAT TO DO
Bumper activated
Press the Safety Lock Button
Program Aborted: Invalid Parameters
Press the Safety Lock Button and try
again
Program Aborted: Incorrect button pressed
Stitching
Press “Move” button
Press Move button for Stitching Initial setup
Press Move button for Stitching Initial setup
Press Move button to extra image position
Press “Move” button
Detector Orientation
Stitching is not possible with Detector in
landscape position. Turn the Detector to
Portrait position.
Tray Out
Insert the Detector Tray.
Tube Rotated and Detector inside
X-Rays inhibited until the Tube is
pointing the Detector. Turn the Tube
towards the Detector.
Emergency Stop
Message associated to a Fatal Error.
Turn the Unit OFF/ON, if error remains,
call field service.