2.7.8
DESIGNATED SIGNIFICANT ZONES OF OCCUPANCY
X-ray equipment specified for examinations that do not need the operator or
staff to be close to the patient during normal use shall be provided with means
to allow the following control functions from a “
Protected Area
” (
refer to
illustration below
):
Selection and control of modes of operation.
Selection of loading factors for the exposure.
Actuation of the exposure controls.
Other necessary controls for the operator during exposure.
ROOM
ELECTRICAL
CABINET
EXAM ROOM
Generator
and
Control Box
POSITIONER
PATIENT ENVIRONMENT
P
A
T
IE
N
T
M
OBI
L
E
TABL
E
OPERATOR CONTROL
PROTECTED AREA
NO RADIATION AREA