DD+DIS150.05E
Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
3
Software Release 1.8.1_C1
Type of new Software*
Software Update
Release Date
10/2006
Printer Serial Numbers
≥
SN 2175 plus 2091, 2134, 2143, 2159
and 2173
Availability
On the Agfa HealthCare Library only the
latest software release can be downloaded.
Prerequisites
None
*NOTE:
A software upgrade is a functional extension. An update (software patch) is an internal
change designed to correct issues, fix defects, or otherwise minor improvements to
performance without changing the software's basic design, structure or functionality.
3.1
Solved Software Issues
Software 1.8.1_C1 solves following software issues:
Symptom 1
Film transport problems from the upper tray
Additional Information:
−
Film jam mainly happens with film size 14x17"
−
Error message HC314003 in the infocounter
−
The user interface shows error message: "Film Jam: Open upper input tray and
clear film path"
Cause
Non-optimized film transport & aligning from upper tray, caused by the friction clutch
in the gear box
Edition 1, Revision 5
DRYSTAR 5302
Chapter 3.8 / 8
02-2015
Type 5366 /100/300