Repair and Service
Machine specific tools, software tools and auxiliary means
DD+DIS150.05E
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 3.2/38
Drystar 5302
Edition 1, Revision 0
Note 2: How to save the " DICOM, LPD and HIPAA report"
In the "Service Engineer Tools" go to the "DICOM, LPD and HIPAA report"
page – right mouse key – save page as.
2.6.1.1
How to view the logfiles offline
Step #
Descpription
Comment
Step 1
Install the PIDEX
tool
Available on MEDNET, GSO Library
Step 2
Extract the
corresponding log
file of the session
file.
See note below
Step 3
In the PIDEX tool
select "Tools –
logview" and open
the desired logfile
Step 4
In the PIDEX tool go
to function "Edit –
find" to find the
desired information.
It is also possible to select a single
word in the logfile. Then all other
occurances of this expression are also
highlighted.
A session file is named e.g.
DS5302_1052_SES_21OCT2005_09-55-48.ZIP, where the different figures
represent following information:
DS5302:
Printer (Drystar 5302)
1052:
Serial Number
SES:
Session File
21OCT2005: Current date
09-55-48:
Current time HH-MM-SS
Interesting for troubleshooting are following files within the SES files:
•
devil.log (controller and print engine);
•
time.log
Agfa Company Confidential
(Type
5366/100)
2005-09-09