Repair and Service
Electrical and mechanical codings, fuse tables and LED’s
DD+DIS150.05E
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 3.4 / 2
Drystar 5302
Edition 1, Revision 0
1.2
Jumper Setting 73 GB Hitachi Hard Disk
536634BC.CDR
1
12
1
24
C3
C4
Figure 2
Jumper C3
Setting
Function
1-2
open
ID Bit 3
3-4
open
ID Bit 2
5-6
open
ID Bit 1
7-8
open
ID Bit 0
9-10
open
Disable Autostart
11-12
open
Force single ended bus mode
13-14
open
Enable delayed start
15-16
open
Write Protect
17-18
open
Reserved
19-20
open
Reserved
21-22
open
LED
23-24
open
Term Power to SCSI bus
Table 3
Jumper C4
Setting
Function
1-2
open
ID Bit 3
3-4
open
ID Bit 2
5-6
open
ID Bit 1
7-8
open
ID Bit 0
9-10
open
Not used
11-12
open
Write Protect
Table 4
Agfa Company Confidential
(Type
5366/100)
2005-09-09