DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 20 of 72
03-2011 Type
5366/100
Agfa Company Confidential
HC304002
: lowerpickupmodule.film.homeposition
Type
Accidental
Context
Action
None
Reason
Lower input jam
Cure
Remove lower input jam
HC304003
: lowerpickupmodule.film.vacuum
Type
Accidental
Context
Action
None
Reason1
Lower input jam
Cure1
Remove lower input jam
Reason2
vacuum
system
failing
Cure2
clean vacuum tubes , filter , vacuum cups ......
Reason3
Lower pickup module defect
Cure3
Replace
Part
vacuum
pump
HC304004
: lowerpickupmodule.filmpickup
Type
Accidental
Context
Action
None
Reason
Lower input jam
Cure
Remove lower input jam