DD+DIS238.06E
Generic Safety Directions
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 5
Generic Safety Directions for HealthCare Imaging Products
Page 25 of 30
19
Safety Directions for Maintenance and Repair Activities
This technical documentation identifies the parts on which preventive inspection and
maintenance shall be performed by Agfa trained service personnel. For required
preventive maintenance frequency refer to the technical documentation or contact
local service management.
In general the device has to be switched off during service activities. Exception: If the
device is switched on to perform tests pay particular attention to any hazards due to
moving and rotating parts. Avoid lose clothing or finger traps. Switch off the device
immediately after the tests.
Do not turn motors manually. If required, first disconnect the motor from the motor
control board.
Make sure that the power cord does not show any signs of damage.
After repair work always check that the integrated safety features are not overridden
or disconnected.
If there is any visible damage to the machine casing do not hand-over the product to
the customer. First repair the machine casing.
Replacing batteries:
WARNING:
Battery can explode, causing chemical burns.
INSTRUCTION:
Check that batteries are inserted with correct polarity.
Only use batteries of the same type or an equivalent type as specified by the
manufacturer.
Dispose of empty batteries in compliance with the specifications of the manufacturer.
When removing batteries from the equipment take appropriate measures to avoid
short circuit of the battery:
Either use tape to cover the two poles of the battery or put the battery back in its
original packing and secure the packing by tape.
04-2014
Agfa Company Confidential