.05E
Repair and Service
Adjustments and Calibrations
DD+DIS150
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
REQUIRED TOOLS:
•
Service PC with network interface and browser Internet Explorer
•
Calibrated densitometer MacBeth TR-924, X-Rite 331, 341, X-Rite 301, X-Rite
310 or Gretag D200-2
•
Lint free cloth
•
Isopropyl - alcohol
•
Waterproof
pen
•
Density reproduction check Excel sheets, available via MEDNET, GSO Library
•
Afssaps / EUREF Calibration Check Excel Sheet, available via MEDNET, GSO
Library
•
Connectivity Release Document for the corresponding modality to DRYSTAR
5302, available via MedNet HealthCare Library
Æ
General Info
Æ
Connectivity &
Application
REQUIRED TIME:
60 to 120 min.
Edition 1, Revision 1
DRYSTAR 5302
Chapter 3.6 / 18
05-2009 Type
5366/100
Agfa Company Confidential