DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 41 of 72
03-2011 Type
5366/100
Agfa Company Confidential
HC307004
: drumdrivemodule.move
Type
Accidental
Context
Action
None
Reason
Drum drive move failed accidently
Cure
Selfresolving
accident
HC307020
: drumdrivemodule.working
Type
Defect
Context
07
Action
Replace
Reason
Drum drive module defect
Cure
Replace
Part
MOD07
HC307021
: drumdrivemodule.m1.shortcircuit
Type
Defect
Context
E307021
Action
SeeManual
Reason
Drum drive module defect
Cure2
Replace
Part
MOD07